Quzhou Aoyin Metal Materials Co.,Ltd

AOYIN Insights

How to Read a Leak Test Report for Battery Trays

How to Read a Leak Test Report for Battery Trays



Annotated leak test report for an aluminium battery tray showing method, pressure, stabilization, acceptance criterion and serial number fields

The report arrives after the boxes do

The leak test report is the only document that tells you whether the enclosure you are about to build into a pack is actually sealed — and it is usually the one nobody reads. A shipment of 300 battery trays lands with a packing list, a certificate of conformity and a one-page test report. The buyer checks the quantity, signs for the pallets, and the report goes into a folder. Six weeks later, three trays come back from the field with water inside.

The failure was already visible in that report. It said "leak test: passed" without naming the method, without a leak rate, and without a serial number tying the result to any tray. Nothing in it was false; none of it was checkable.

Why a bad report costs more than no report

A report you trust wrongly is worse than no report, because it closes the question. Price the two directions. A tray that leaks in service takes the pack with it: cells, BMS and harness, plus the labor to pull the pack off a vehicle already sold. Under AIS 156 Phase 2, lithium traction packs on L-category vehicles in India have had to meet IPx7 since 31 March 2023 — one meter of immersion for 30 minutes per IEC 60529, as set out in IP54 vs IP65 vs IP67 — so a sealing failure is also a compliance question.

The other direction is cheap: asking for four more fields costs an email, and re-testing 5–10 parts per lot at an outside laboratory costs a few hundred dollars. Both are trivial against one field failure, which is why the checklist below belongs with the enquiry and not with the claim.

The 9 fields a real leak test report carries

A credible report names the method, the pressure, the stabilization, the acceptance criterion and the unit it applies to — the other four fields are what let you audit it later. Anything missing from this list is a gap you are being asked to fill with trust.

Report fieldWhat a credible value looks likeImpact on your acceptance
Test methodnamed: pressure decay, differential pressure, or helium vacuum chamber per ISO 20485The method sets the resolution; without it the number means nothing
Test pressure10 kPa for pressure screens; 5–10 kPa for helium sniffingWrong pressure invalidates the result on a vented enclosure
Stabilization time≥30 s, statedSkipping it converts thermal drift into false passes
Test duration180 s typical for differential pressureShort tests under-detect slow leaks
Acceptance criterion≤50 Pa drop (screen); ≤1×10⁻⁶ Pa·m³/s (helium)A report without a criterion cannot fail anything
Instrument resolution10⁻⁴ Pa·m³/s for differential pressureTells you the smallest leak the station could see
Temperature at testrecorded, within ±1 °CA 1 °C shift moves a 100-litre reading by roughly 0.4%
Calibration datereference leak calibrated within 12 monthsAn uncalibrated instrument produces confident, meaningless numbers
Unit identificationserial number or lot ID per resultBatch-level "pass" cannot be traced to a failed tray

Two fields do most of the work. Resolution tells you what the station was physically capable of detecting; the acceptance criterion tells you what the supplier actually promised — a report can honestly pass a tray against a criterion far looser than the one you assumed. The distinction between what production screening can see and what a type test can measure is set out in how battery boxes are leak tested, and the method set behind it in the sealing and leak testing hub.

Which report your volume actually needs

Match the report to the quantity, not to the most impressive method on the list. Asking a 50-piece pilot run for full helium documentation adds cost without adding assurance; accepting a visual check on a 5,000-piece program is where the money goes.

  • 1–10 sets (prototype): helium vacuum-chamber report per serial number. The enclosure design is still changing, and the type-test number is what tells you the design works. Cost is irrelevant at this quantity.

  • 50–500 sets (pilot): differential-pressure screening on every unit, plus helium verification on 5–10 parts per lot at an independent laboratory. The screened report proves consistency; the sampled helium report proves the screen is calibrated against reality.

  • 5,000+ sets (volume): line-side differential pressure at 100%, first-article helium type test, then periodic sampling with a stated cadence. Unless your annual volume justifies the chamber, the rational setup is screening in-house plus outside verification at an independent laboratory.

The mistake in both directions is the same: asking one document to do another's job. The report should say which test it is.

Decision path for battery tray leak test reports by order quantity: prototype helium, pilot screening plus sampled helium, volume screening plus first-article type test

How a passing report hides a leaking box

Most false passes are not falsified results — they are real measurements read against the wrong criterion. Six patterns account for nearly all of them:

What the report showsRoot causePrevention, with the number
"IP67 compliant", no leak ratethe test was a visual or immersion checkrequire a leak rate in Pa·m³/s with the method named
One "pass" for the whole batchresults not captured per unitrequire a result row per serial number
No stabilization time statedthermal drift read as seal performancerequire ≥30 s stabilization and the test temperature recorded
Differential pressure claiming 10⁻⁶resolution quoted from the wrong instrumentdifferential pressure resolves 10⁻⁴; 10⁻⁶ belongs to helium
Pass at 08:00, fail at 17:00test cell next to a loading dockrequire the test cell temperature band, ±1 °C
Calibration date over a year oldreference leak never re-certifiedrequire calibration within 12 months, with the certificate number

The third one deserves emphasis, because it is invisible in the finished number. A 30 s stabilization window against a 180 s test is about 15% of the cycle, and cutting it is the easiest way to raise throughput. The report looks identical either way.

The 9-point checklist before you accept the shipment

  1. Test method named — pressure decay, differential pressure, or helium per ISO 20485

  2. Test pressure stated in kPa, and whether the pressure-equalisation vent was blocked

  3. Stabilization time ≥30 s, written on the report

  4. Acceptance criterion with a number: Pa drop or Pa·m³/s

  5. Instrument resolution stated, so you know the detection floor

  6. Test temperature recorded, within ±1 °C

  7. Calibration date within 12 months, with the reference-leak certificate number

  8. A result row per serial number, not a batch statement

  9. Sample re-test clause in the purchase order: 5–10 parts per lot, independent laboratory

Send the nine fields with the enquiry, not after delivery. A supplier running the test already has all nine; a trader has to go and ask someone.

Red flags in a leak test report

  • "IP67 passed" with no leak rate. IP67 is a test defined by IEC 60529, not a number. A report quoting only the code was written, not measured.

  • No equipment identity. No method, no instrument, no serial — nothing to audit six months later.

  • One report reused across lots. Same file number, same date, three shipments. The test was run once, if at all.

  • A criterion looser than your drawing. The report is honest and still worthless: it tells you the box passed their number.

  • No answer when you ask which test each finished tray passes. If the name of the method cannot be produced in one email, the method is not running.

  • Certificates without an issuing body, or a glossy brochure in place of a dimensioned sample report.

Differential pressure screening and helium type testing

Every unit gets screened on the line; first articles and sampled lots get the helium number.

Inspector checking a machined aluminum battery tray part against specification before leak testing

Aluminium battery enclosures for e-rickshaw and e-scooter packs run through an integrated production network: extrusion up to 5,000 t presses, CNC machining, friction stir welding, anodizing and leak testing under one quality system. Screening is differential pressure at 10 kPa on every unit; type testing is vacuum-chamber helium at ≤1×10⁻⁶ Pa·m³/s on first articles and sampled lots, referenced to ISO 20485. Reports carry all nine fields above, and audit is welcome at every stage.

Frequently asked questions

What leak rate should I specify for an IP67 battery tray? There is no single conversion, because IP67 is a test and not a leak rate. In practice a tray that passes IP67 typically measures in the 10⁻⁶ Pa·m³/s range or better by helium, so specifying ≤1×10⁻⁶ Pa·m³/s with the method named is the defensible line.

Can a differential-pressure report prove IP67? Not on its own. Differential pressure resolves 10⁻⁴ Pa·m³/s, which catches any defect that actually occurs in production but will not resolve a 10⁻⁶ leak. Use it as the 100% screen and require helium on first articles and sampled lots.

How many parts should be re-tested independently? 5–10 per lot for pilot and early volume. Below about 1,000 parts a year, buying a helium chamber does not pay back, so outside verification is the rational route.

Should the pressure-equalisation vent be blocked during the test? The report should say which was done. The vent is designed to pass air while blocking water, so either the test pressure is chosen to keep vent flow below the acceptance threshold, or the vent is sealed for the test — and the report must state which.

What if the report has no serial numbers? Treat it as a batch statement, not a test record. Ask for a result row per unit; if the station does not capture serial-level data, it cannot support a claim about a specific failed tray.

Send the nine fields with your enquiry to info@aymetals.com, or reach us on WhatsApp at +86 133 0570 9557 — the same eight specification items that fix the drawing — or ask for a sample report before you place the order.

Related guides

Diagram: Aoyin Metals schematics. Acceptance figures cite ISO 20485 for tracer-gas testing and IEC 60529 for IP codes.

Need Aluminum Battery Enclosures?

From extrusion to CNC machining and anodizing, AOYIN delivers end-to-end manufacturing for EV battery housings.